CNVLOSS Conversion Loss

Syntax
Figure 7-80
CNVLOSS Syntax
Description
The CNVLOSS command compensates for losses outside the
instrument when in external mixer mode (such as losses within
external mixers or connector cables). CNVLOSS specifies the mean
conversion loss for the current harmonic band. In a full frequency band
(such as band K), the mean conversion loss is defined as the minimum
loss plus the maximum loss for that band divided by two. Adjusting for
conversion loss allows the system to remain calibrated (that is, the
displayed amplitude values have the conversion loss incorporated into
them). The default value for any band is 30 dB. The spectrum analyzer
must be in external-mixer mode in order for this command to work.
When in internal-mixer mode, querying CNVLOSS returns a zero. This
function is not available for an 8560E/EC Option 002.
Parameters
number
UP or DN
Preset State
30 dB
Chapter 7
real from 15 to 60.
0.1 dB.
Language Reference
CNVLOSS Conversion Loss