Aunits Absolute Amplitude Units; Figure 7-62 Aunits Syntax - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Language Reference

AUNITS Absolute Amplitude Units

AUNITS Absolute Amplitude Units
Syntax
Figure 7-62
AUNITS Syntax
Description
The AUNITS command sets the absolute amplitude units for the input
signal and the display. AUNITS will affect the query responses of the
following commands: MKA, TRA and TRB (when in trace data format
P-format), DL, RL, SQUELCH, TH, and VTL. AUNITS is disabled
when the 8560E/EC Option 002 tracking generator is in use.
Parameters
AUTO
MAN
Preset State
Coupled mode, dBm
426
sets amplitude units to coupled mode. For a log scale,
the units default to dBm; for a linear scale, units
default to volts.
sets amplitude units to manual mode.
Chapter 7