The measurement method and the number of alternate channels you
have selected determine the size of the data array that will be returned
by the ACPRSLTS command. Table 7-1 indicates the values that will be
returned for each method. The values are returned in the order
indicated in the table. One set of values is returned if the number of
alternate channels is set to zero. Each additional alternate channel pair
selected will return another set of values. See Table 7-4 on page 402.
Table 7-3
ACP Methods Versus Results (with 0 Alternate Channels)
Method
Analog
Peak
2BW
Gated
Burstpwr
Chapter 7
ACPRSLTS Adjacent Channel Power Measurement Results
Number of
Results (in order of output)
Results/Set
2
ACP ratio (lower channel)
ACP ratio (upper channel)
2
ACP ratio (lower channel)
ACP ratio (upper channel)
6
Random power (lower channel)
Impulsive power (lower channel)
Weighted power (lower channel)
Random power (upper channel)
Impulsive power (upper channel)
Weighted power (upper channel)
6
Random (lower channel)
Impulsive power (lower channel)
Weighted power (lower channel)
Random (upper channel)
Impulsive power (upper channel)
Weighted power (upper channel)
8
Random power (lower channel)
Impulsive power (lower channel)
Weighted power (lower channel)
Average power (lower channel)
Random power (upper channel)
Impulsive power (upper channel)
Weighted power (upper channel)
Average power (upper channel)