Rbr Resolution Bandwidth To Span Ratio; Figure 7-219 Rbr Syntax; Figure 7-220 Rbr Query Response - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

RBR Resolution Bandwidth to Span Ratio

Syntax
Figure 7-219
RBR Syntax
Description
The RBR command specifies the coupling ratio between the resolution
bandwidth and the frequency span. When the frequency span is
changed, the resolution bandwidth is changed to satisfy the selected
ratio.
Parameters
number
UP or DN
Preset State
0.011
Query Response
Figure 7-220
RBR Query Response
Chapter 7
real from 0.002 to 0.10.
increments in a 1, 2, 5 sequence.
Language Reference
RBR Resolution Bandwidth to Span Ratio
571