Making Measurements
Example 4: Harmonic Distortion
An Alternative Harmonic Measurement Method:
Procedure B
This method is somewhat longer, but because each signal is measured
in a narrower span and resolution bandwidth, the signal-to-noise ratio
is improved, making the results more accurate.
1. Using the present setup, clear the markers from the screen by
pressing
,
MKR
MARKERS OFF
is pressed, the display expands to the full size of the screen, for
HOLD
easier viewing.
Measure the 1 MHz fundamental
1. Reduce the frequency span:
a. Press
PEAK SEARCH
highest-level signal on-screen.
b. Reduce the frequency span to 1 MHz by pressing
.
ZOOM
c. Enter 1 MHz. The span zoom function lets you quickly "zoom" to a
narrower frequency span without losing the signal from the
screen (because signal track is activated).
d. After the frequency span is reduced, turn off the signal-track
function by pressing
2. Set the center-frequency step-size to the frequency of the
fundamental:
MKR →
MARKER → CF STEP
a. Press
,
3. For better amplitude accuracy, move the signal to the top of the
graticule:
MARKER → REF LVL
a. Press
Figure 2-19 on page 73.
72
. Notice that when
, which activates a marker on the
, and setting
MKR
SIG TRK ON OFF
. The resulting display should resemble
or
MARKERS OFF
,
SPAN
SPAN
to OFF.
Chapter 2