HNUNLK Unlock Harmonic Number

Syntax
Figure 7-137
HNUNLK Syntax
Description
The HNUNLK command unlocks the harmonic number, allowing you to
select frequencies and spans outside the range of the locked harmonic
number. Also, when HNUNLK is executed, more than one harmonic can
then be used to sweep across a desired span. For example, sweep a span
from 18 GHz to 40 GHz. In this case, the analyzer will automatically
sweep first using 6− mixing harmonic, then using 8− mixing harmonic.
This command is not available with an 8560E/EC Option 002.
Example
10
20
30
40
Chapter 7
OUTPUT 718;"IP;MXRMODE EXT;FULBAND Q;"
OUTPUT 718;"FA 18GHZ;FB 40GHZ;"
OUTPUT 718;"HNUNLK;"
END