Mkbw Marker Bandwidth; Figure 7-156 Mkbw Syntax - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Language Reference

MKBW Marker Bandwidth

MKBW Marker Bandwidth
Syntax
Figure 7-156
MKBW Syntax
Description
When used remotely, the MKBW command finds the signal bandwidth
at the power level in dB below the on-screen marker (if a marker is
present) or the signal peak (if no on-screen marker is present). When
the command is used manually, a peak search is automatically
performed, and the bandwidth of the largest signal on-screen is
displayed in the message area.
Parameter
integer
Example
10
20
30
40
50
60
508
any valid negative integer (0 to the amplitude of the
noise floor).
OUTPUT 718;"IP;"
OUTPUT 718;"CF 300MHZ;SP 100MHZ;SNGLS;"
OUTPUT 718;"TS;MKPK HI;"
OUTPUT 718;"MKBW -3,?;"
ENTER 718;Mkbw
END