ACPALTCH Adjacent Channel Power
Alternate Channels
Syntax
Figure 7-8
ACPALTCH Syntax
Description
The ACPALTCH command sets the number of alternate channels to be
measured by an adjacent channel power measurement to either 0, 1, or
2. Zero is the default. The number of alternate channels is used with
the ACPRSLTS command.
Query Response
Figure 7-9
ACPALTCH Query Response
Example
10
20
30
40
Chapter 7
INTEGER Altchannel
Altchannel = 2
OUTPUT 718;"ACPALTCH";Altchannel;";"
END