SAVES Save State

Syntax
Figure 7-234
SAVES Syntax
Description
The SAVES command saves the currently displayed instrument state in
the specified state register.
Parameters
number
PWRON
Example
10
20
30
Chapter 7
integer from 0 to 9. Registers 8 and 9 are used by the
normalization routine. Numbers less than zero default
to zero; numbers greater than nine default to nine.
sets the spectrum analyzer to the current state when
the spectrum analyzer is turned on.
OUTPUT 718;"IP;CF 300MHZ;SP 20MHZ;"
OUTPUT 718;"SAVES PWRON;"
END