This manual uses the following conventions:
This font represents either a hard key, which is physically
Front-Panel Key
located on the instrument, or a softkey, whose label is
determined by the instrument's firmware.
This font indicates text displayed on the instrument's
Screen Text
screen.
Differences between 8560 EC-Series and
E-Series Spectrum Analyzers
Features
8560 EC-Series
Display
— LCD display
— color
— display not adjustable
— backlight bulbs are
replacable (replace both
bulbs when display is dim)
— requires A17 LCD driver
board
Fast ADC
— FADC is standard
sweep times
— FADC circuitry integrated
(FADC)
into A2 controller board
enables sweep
times of 30 ms
to 50 µs
VGA port
— located on rear panel
— always active
— does not require user
interface
In all other operational respects the EC-series and E-series are
identical. Unless otherwise noted, the information in this manual
applies to all EC-series and E-series instruments.
Chapter 1
Quick Start Guide
What You'll Find in This Chapter
8560 E-Series
— CRT display
— monochrome
— display adjustable for
focus, intensity, and
quadrature
— requires high voltage
module (HVM), which is
located in the A6 power
supply
— requires A17 CRT driver
board
— FADC is available as an
option (Option 007)
— separate A16 FADC board
required
Not available