Agilent Technologies 8560E User Manual page 58

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Making Measurements
Example 2: Improving Amplitude Measurements with Ampcor
6. On the spectrum analyzer, press
and
EDIT AMPCOR
pressing
MORE 1 OF 2
you can save the correction first and then purge it. Refer to the
procedures that follow.) If data was purged, press
again before continuing.
Enter the correction points.
1. On the spectrum analyzer, press
.
2 OF 2
2. Enter the marker frequency.
3. Use the knob to adjust AMPL until the marker amplitude reads the
same as the power meter. The ampcor function will be automatically
turned on after the first point is entered. Note the "W" along the left
edge of the display.
4. Set the source CW frequency to the next frequency to be corrected.
5. Enter the power sensor calibration factor for this frequency into the
power meter.
6. On the spectrum analyzer, press
frequency to be corrected.
7. Press
,
CAL
MORE 1 OF 2
,
,
2
PEAK SEARCH
8. Use the knob to adjust AMPL until the marker amplitude reads the
same as the power meter.
9. Repeat steps 4 through 8 for each additional correction point.
Storing the ampcor data.
1. Press
MORE 1 OF 2
register number.
2. If you would like to create a softkey label under which the correction
data will be saved, press
the ampcor menu, press
register number. The first 16 characters of the screen title will be
used as a softkey label.
You have now completed the process for entering amplitude correction
points. At this point you can disconnect the source from the spectrum
analyzer input and connect the test signal, using the same system
cable. You can recall the correction points and begin to make
measurements.
58
,
CAL
MORE 1 OF 2
. If there is a correction already loaded, purge it by
,
,
DONE EDIT
PURGE CORR
MORE 1 OF 2
FREQUENCY
,
AMPCOR MENU
,
MORE 2 OF 2
NEW CORR PT
,
,
DONE EDIT
SAVE AMPCOR
and enter a screen title. Return to
DISPLAY
SAVE AMPCOR
,
AMPCOR MENU
,
PURGE DATA
EDIT AMPCOR
,
PEAK SEARCH
and enter the next
,
,
EDIT AMPCOR
MORE 1 OF
.
. Select a convenient
and select a correction
Chapter 2
,
. (Or
,
MORE

Advertisement

Table of Contents
loading

Table of Contents