Apb Trace A Plus Trace B; Figure 7-59 Apb Syntax - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

APB Trace A Plus Trace B

Syntax
Figure 7-59
APB Syntax
Description
The APB command adds the contents of trace A to trace B and stores
the result in dBm (when in log mode), in trace A. When in linear mode,
the results are in volts. Trace A is placed in view mode. This command
is done immediately and not on a repetitive basis.
The displayed amplitude of each trace element falls in one of 600 data
NOTE
points. There are 10 points of overrange, which corresponds to one-sixth
of a division of overrange. When adding or subtracting trace data, any
results exceeding this limit are clipped at the limit.
Example
10
20
30
40
Chapter 7
OUTPUT 718;"IP;SNGLS;"
OUTPUT 718;"TS;VIEW TRA;CLRW TRB;TS;VIEW TRB;"
OUTPUT 718;"APB;BLANK TRB;"
END
Language Reference
APB Trace A Plus Trace B
423