Pwrbw Power Bandwidth (Full Trace); Figure 7-215 Pwrbw Syntax; Figure 7-216 Pwrbw Query Response - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

PWRBW Power Bandwidth (Full Trace)

Syntax
Figure 7-215
PWRBW Syntax
Description
The PWRBW command first computes the combined power of all signal
responses contained in a trace array. The command then computes the
bandwidth equal to a percentage of the total power.
For example, if 100 percent is specified, the power bandwidth equals
the current frequency span. If 50 percent is specified, trace elements
are eliminated from either end of the array, until the combined power of
the remaining trace elements equals half of the total power computed.
The frequency span of these remaining trace elements is the power
bandwidth output to the controller.
Parameter
number
Query Response
Figure 7-216
PWRBW Query Response
Chapter 7
real from 0.00 to 100.00.
Language Reference
PWRBW Power Bandwidth (Full Trace)