Figure 2-47 . Signal Responses Produced By A 50 Ghz Signal In U Band - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Figure 2-47
Signal Responses Produced by a 50 GHz Signal in U Band
Identify signals with the frequency-shift method
6. Signal-identification routines that identify the signal and images are
Chapter 2
available on instruments with firmware revisions ≤920528, or with
Option 008. The frequency-shift method of identifying valid signals
uses the spectrum-analyzer function
function, smaller spans will yield more accurate measurements.
a. Span down to 50 MHz.
b. Press
,
AUX CTRL
EXTERNAL MIXER
c. Press
SIG ID ON OFF
Any signal not produced by the currently selected harmonic will be
shifted horizontally on alternate sweeps (see Figure 2-48).
Therefore, when viewing the display, on alternate sweeps the signal
will appear, then not appear, at the frequency of interest. The correct
signal produced by the selected harmonic will be shifted in a vertical
direction only, as in Figure 2-49.
Example 8: External Millimeter Mixers (Unpreselected)
SIG ID ON OFF
,
SIGNAL IDENT
until ON is selected.
Making Measurements
. When using this
.