RCLTHRU Recall Thru

Syntax
Figure 7-224
RCLTHRU Syntax
Description
The RCLTHRU command recalls the internally stored thru-reference
trace into trace B. The instrument state is also set to the stored
thru-reference state.
Example
The following example applies only to an 8560E/EC Option 002.
However, the RCLTHRU command can be used on the 8560 E-Series and
EC-Series spectrum analyzers when they are being used with an Agilent
85640A tracking generator, an Agilent 85644A tracking source, or
Agilent 85645A tracking source.
10
20
30
40
50
60
70
RE."
80
90
100
110
120
130
140
150
160
Chapter 7
OUTPUT 718;"IP;SNGLS;"
OUTPUT 718;"FA 300 KHZ;FB 1GHZ;"
OUTPUT 718;"SRCPWR ON;"
OUTPUT 718;"SWPCPL SR;"
OUTPUT 718;"SRCTKPK;DONE?;"
OUTPUT 718;"ENTER 718;Done
PRINT "CONNECT THRU.
PAUSE
OUTPUT 718;"TS;DONE?;"
ENTER 718;Done
OUTPUT 718;"STORETHRU;TS;DONE?;"
ENTER 718;Done
PRINT "RECONNECT DUT.
PAUSE
OUTPUT 718;"NORMLIZE ON;"
OUTPUT 718;"TS;DONE?;"
PRESS CONTINUE WHEN READY TO STO
PRESS CONTINUE WHEN READY."