Replacing The Battery - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Replacing the Battery

If the battery that maintains the spectrum analyzer random access
memory is deteriorating, stored states and traces will only be retained
for a short time after the instrument is powered off. When the voltage
across the battery drops to +2.6 V, its life and use are limited and the
output voltage will deteriorate quickly. The battery for maintaining the
memory can be accessed through the rear panel.
There are two ways to avoid loss of the calibration constants.
• Remove and replace the battery in less than 10 minutes.
• Power on the spectrum analyzer before removing the battery.
Replace the battery using the following steps:
1. If an option module is attached to the rear panel, turn the spectrum
2. Remove the two screws securing the small panel labeled with a
3. Remove the battery and replace it, observing proper polarity. Use
Battery part number
Voltage
Amps
Typical life
Type
The battery contains lithium polycarbon monofluoride. Do not
WARNING
incinerate or puncture this battery. Dispose of discharged
battery in a safe manner.
Chapter 9
analyzer off and remove the module. Then, if necessary, turn the
spectrum analyzer on.
battery. Do not remove the bottom screw.
the following Agilent Technologies battery, part number:
Battery Information
1420-0341
3 V
1.2 A
3 to 5 years
non-rechargeable
If You Have a Problem
Replacing the Battery
671

Advertisement

Table of Contents
loading

Table of Contents