Agilent Technologies 8560E User Manual page 690

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

If You Have a Problem
Error Messages
Error Code Listing
Error codes and their associated messages are listed in numeric order
below. Error codes 100 to 199 relate to incorrect spectrum
analyzer programming
NO PWRON Power-on state is invalid; default state is
ERR 100
loaded.
NO STATE State to be RECALLed not valid or not SAVEd.
ERR 101
# ARGMTS Command does not have enough arguments.
ERR 102
# ARGMTS Command does not have enough arguments.
ERR 103
# ARGMTS Command does not have enough arguments.
ERR 104
# ARGMTS Command does not have enough arguments.
ERR 105
ABORTED! Current operation is aborted; GPIB parser
ERR 106
reset.
HELLO ?? No GPIB listener is present.
ERR 107
TIME OUT Analyzer timed out when acting as controller.
ERR 108
CtrlFail Analyzer unable to take control of the bus.
ERR 109
NOT CTRL Analyzer is not system controller.
ERR 110
# ARGMTS Command does not have enough arguments.
ERR 111
??CMD?? Unrecognized command.
ERR 112
FREQ NO! Command cannot have frequency units.
ERR 113
TIME NO! Command cannot have time units.
ERR 114
AMPL NO! Command cannot have amplitude units.
ERR 115
?UNITS?? Unrecognizable units.
ERR 116
NOP NUM Command cannot have numeric units.
ERR 117
NOP EP Enable parameter cannot be used.
ERR 118
NOP UPDN UP/DN are not valid arguments for this
ERR 119
command.
NOP ONOF ON/OFF are not valid arguments for this
ERR 120
command.
NOP ARG AUTO/MAN are not valid arguments for this
ERR 121
command.
NOP TRC Trace registers are not valid for this command.
ERR 122
NOP ABLK A-block format not valid here.
ERR 123
690