Language Reference

STORETHRU Store Thru

STORETHRU Store Thru
Syntax
Figure 7-269
STORETHRU Syntax
Description
The STORETHRU command stores a thru-calibration trace into trace B
and into the nonvolatile memory of the spectrum analyzer.
The state of the thru information is stored in state register number 9.
The STORETHRU command is primarily intended for use with a
NOTE
tracking generator.
A normalization error will occur when the NRL command is executed if
the current spectrum analyzer state is not the same as the state stored
by the last execution of the STORETHRU or STORESHORT command.
This can happen when several open/short or thru calibrations are
performed.
The NRL command recalls the last calibration run. If one of the earlier
short or thru states has been recalled just before executing the NRL
command, then the current and recalled states don't match and the
error occurs. To avoid this error, update the STORETHRU or
STORESHORT state register with the current state before turning
NRL on. The STORETHRU state register is state register 9. The
STORESHORT state register is state register 8. It is a good idea to take
a sweep (TS command) prior to saving the state using the SAVES 8 or
SAVES 9 command.
624
Chapter 7