Source Level Accuracy And Source Power Linearity Test (Osc Test); Speci Cation; Required Equipment; Procedure - Agilent Technologies E4915A Service Manual

Crystal impedance / lcr meter
Table of Contents

Advertisement

3. Source Level Accuracy and Source Power Linearity Test
(OSC TEST)
Note
E4915A does not require performing the Source Power Linearity Test. The
program will skip the Source Power Linearity Test for the E4915A.
Source Level Accuracy and Source Power Linearity Test are
performance tests. The test data of those tests are saved in the
C:\E4915_16\DATA_15A\SNxxxxx or C:\E4915_16\DATA_16A\SNxxxx le.
Speci cation
Level Accuracy:
Power Lineality

Required Equipment

Power Meter
Power Sensor
N(f)-BNC(m) Adapter

Procedure

1. Connect the test equipment as shown in Figure 2-5.
Figure 2-5. Source Level Accuracy and Source Power Linearity Test Setup
6
2 dB (10 MHz or 50 MHz, 5 dBm
output level)
6
3 dB (1 MHz to 180 MHz, 5 dBm
output level)
6
2 dB (relative to 5 dBm output
level at 10 MHz)
436A, 437B, or 438A
8482A
PN 1250-1477
Adjustments and Performance Tests 2-11