Figure 2-50 Sig Id At Mkr Performed On An Image Signal - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Identify signals in wide frequency spans
7.
8. To exit the external mixer mode, press
Figure 2-50
SIG ID AT MKR
Chapter 2
identifies signals in wide frequency spans, using
SIG ID AT MKR
harmonic search.
SIG ID AT MKR
frequency of a signal and displays its value on the spectrum
analyzer.
Activating
SIG ID AT MKR
reading in the active block, as shown in Figure 2-50. The true signal
frequency is given with an identifier that the marker resides on an
image. If the marker is placed on a true signal, and
activated, the signal frequency will appear without the IMAGE
notation, as shown in Figure 2-51. To activate
a. Place a marker on a signal.
b. Press
,
AUX CTRL
EXTERNAL MIXER
.
MIXER
Performed on an Image Signal
Example 8: External Millimeter Mixers (Unpreselected)
automatically determines the proper
on an image of the signal will yield a
,
SIGNAL IDENT
AUX CTRL
Making Measurements
is
SIG ID AT MKR
:
SIG ID AT MKR
,
SIG ID AT MKR
, then
INTERNAL
105
.