Agilent Technologies 8560E User Manual page 227

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

The counted value appears in the upper right corner of
the display.
Front-panel key access:
COUNTER
Adjusts the resolution of the frequency-count
RES
measurement. The resolution ranges from 1 Hz to 1
MHz in decade increments. The default value is 10
kHz. The counter measurement occurs over a time
interval of twice the reciprocal of the counter resolution
for resolution bandwidths greater than or equal to 300
Hz.
Front-panel key access:
COUPLING
AC DC
Not available with an Agilent 8563E/EC, Agilent
8564E/EC, or Agilent 8565E/EC. Toggles the input
between ac and dc coupling. The preset state of the
analyzer is ac-coupled. This protects the input of the
analyzer from damaging dc signals. It also limits the
lower frequency-range response to 100 kHz (although
the analyzer will tune down to 0 Hz with signal
attenuation).
Front-panel key access:
CRT ADJ
Displays a fixed pattern of lines and characters on the
PATTERN
screen, each of which is used in setting the various
adjustments in display hardware (such as vertical gain,
focus, and so on) that are available in E-series
instruments. Three of these adjustments—X POSN, Y
POSN, and TRACE ALIGN—are available from the
rear panel. Use the CRT alignment pattern and the
above three adjustments to align the display. Refer to
Figure 4-3. When this function is active, all front-panel
keys are inactive except
the pattern and return the analyzer to its previous
state. For other display adjustments, refer to the
Chapter 8. EC-series instruments do not require
adjustment and are not adjustable.
Front-panel key access:
Chapter 4
Key Function Descriptions
Key Descriptions
FREQ COUNT
FREQ COUNT
AMPLITUDE
. Press EXIT to remove
PRESET
CAL
227