Figure 2-42 Norm Ref Lvl Is A Trace Function; Figure 2-43 Range Lvl Adjusts Analyzer For Compression-Free Measurements - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Making Measurements
Example 7: Stimulus-Response Measurements
Figure 2-42
NORM REF LVL
After returning
shown in Figure 2-43, the trace moves fully within the graticule.
Compare the settings: (1) input attenuator value has changed to 40 dB,
(2) the marker-amplitude readout displays −6.3 dB, and (3) the ERR 903
A>DLMT error message no longer appears.
Figure 2-43
RANGE LVL
96
is a trace function.
to 0 dB, increase
NORM REF LVL
adjusts analyzer for compression-free measurements.
to 30 dB. As
RANGE LVL
Chapter 2