Procedures

The following adjustment and troubleshooting procedures are included
here.:
• Trace Alignment - used for 8560 E-Series instruments only
• Reference Level Calibration
• GPIB Address Selection
• Plotting And Printing Directly
Trace Alignment (8560 E-series)
1. Press
2. Adjust the rear-panel TRACE ALIGN until the leftmost line of the test
3. Adjust the rear-panel X POSN until the leftmost @ characters and the
4. Adjust the rear-panel Y POSN until the softkey labels align with the
5. Press
Figure 9-3
CRT Adjustment Pattern
Chapter 9
PRESET CAL, MORE 1 OF 2
pattern is parallel with the CRT bezel. See Figure 9-3 on page 675.
softkey labels appear just inside the left and right edges of the CRT
bezel.
appropriate softkeys.
to return the analyzer to normal operation.
PRESET
If You Have a Problem
,
.
CRT ADJ PATTERN
Procedures
675