Page 2
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent...
Page 3
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For assistance, refer to “Contacting Agilent” on page...
Page 4
Contacting Agilent Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact Agilent for assistance. Online assistance: www.agilent.com/find/assist United States Latin America Canada Europe (tel) 1 800 452 4844, or (tel) (305) 269 7500 (tel) 1 877 894 4414 (tel) (+31) 20 547 2323 1 800 593 6635 for on-site (fax) (305) 269 7599...
Page 5
Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards. The WARNING notice denotes a hazard. It calls attention to a WARNING procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury.
Page 6
Safety Earth Ground This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
Page 7
Typeface Conventions Italics • Used to emphasize important information: Use this software only with the Agilent Technologies xxxxxX system. • Used for the title of a publication: Refer to the Agilent Technologies xxxxxX System-Level User’s Guide. • Used to indicate a variable: Type LOAD BIN filename.
Page 8
Using this document and an Agilent Technologies 85103E upgrade package, a trained Agilent Technologies service engineer can upgrade an Agilent Technologies 8510A (with serial number below 03000) to full Agilent Technologies 8510C capability. Agilent Technologies does not recommend customer installation.
• Check the 8510C operation. • Load and check the calibration kit definitions. Agilent Technologies does not recommend customer installation of this NOTE upgrade. The Agilent 85103E performance upgrade package comes with prepaid installation. Refer to “Contacting Agilent” on page iv to arrange for a service engineer to install the upgrade package.
85103E Upgrade Package (8510A to 8510C) Installing the Upgrade Performance Verification Performance verification of the system is not included in the upgrade package. If you wish to do a performance verification of your upgraded system, this additional service can be arranged with the Agilent service engineer installing your upgrade.
85103E Upgrade Package (8510A to 8510C) Installing the Upgrade Procedure Overview Step 1. Check the upgrade package contents. Refer to Table 1. If an item is missing, refer to “Contacting Agilent” on page iv assistance. Step 2. Assemble items not included in the upgrade package.
Page 15
85103E Upgrade Package (8510A to 8510C) Installing the Upgrade Table 1 85103E Upgrade Package Contents Item Quantity Part Number Assemblies A16 remote applications 85102-60235 A20 sweep ADC 85102-60234 A23 main phase lock 85102-60240 85102 emulator 85101-60209 Tools 0.655-inch countersink drill bit...
Page 16
85103E Upgrade Package (8510A to 8510C) Installing the Upgrade Table 1 85103E Upgrade Package Contents (Continued) Item Quantity Part Number Documentation Upgrade Installation Manual 85103-90013 p/o 85101C 8510C Manual set contains: 08510-90275 8510C Keyword Dictionary 08510-90280 8510C Operating and Programming Manual...
Page 17
85103E Upgrade Package (8510A to 8510C) Installing the Upgrade Table 2 Equipment Required But Not Supplied Item Agilent Part Number Source 8360-series Test Set 8511A/B, 8514B, 8515A, 8517A/B, or 85110A/L Static control table mat 9300-0797 Wrist strap 9300-1367 Wrist-strap-to-mat cord (5 ft)
85103E Upgrade Package (8510A to 8510C) Checking 8510 System Operation Checking 8510 System Operation Use this procedure to ensure that the 8510A system works properly. Procedure 1. Configure the system as shown in Figure 2. Turn on the instruments in the following order: a.
Page 19
85103E Upgrade Package (8510A to 8510C) Checking 8510 System Operation Figure 1 Typical Equipment Setup and 8511A-Specific Connections 85103E Performance Upgrade Installation Manual...
85103E Upgrade Package (8510A to 8510C) Converting Files from Tape to Disk Converting Files from Tape to Disk Each performance upgrade package includes a tape to disk conversion program. Use this program to convert DATA, MEMORY, MEMORY ALL, and DELAY TABLE files from tape to disk.
85103E Upgrade Package (8510A to 8510C) Converting Files from Tape to Disk 5. On the controller, load the program: a. Type GET “TAPE_DISC”, and press Return b. Wait approximately 2 minutes. NOTE You can store a version of this program so that it loads faster for future use.
85103E Upgrade Package (8510A to 8510C) Converting Files from Tape to Disk Storing a Binary Version To store a binary version of the utility program for faster loading in the future: 1. Type STORE "TAPEDISC". 2. To load this program in the future, type LOAD "TAPEDISC".
85103E Upgrade Package (8510A to 8510C) Installing a BNC Connector Installing a BNC Connector Use the following procedure to install a BNC connector on the rear panel of the 85102. WARNING The following procedure presents a danger to your eyes; the law requires that you use safety glasses during such procedures.
Page 24
85103E Upgrade Package (8510A to 8510C) Installing a BNC Connector 5. Remove the alignment guide, and center punch the marked spot (if you do not have an automatic center punch, use the countersink drill bit and a small hammer). 6. Cut the foam tape into four pieces (each three inches long).
Page 25
85103E Upgrade Package (8510A to 8510C) Installing a BNC Connector 12. At the spot punched on the rear panel, use the supplied 0.665-in bit to drill the BNC hole. Keep the following in mind: • Use a constant but slow clockwise RPM.
85103E Upgrade Package (8510A to 8510C) Installing Assemblies Installing Assemblies Use the following procedure to replace three assemblies in the 85102. CAUTION Assemblies in this procedure are very sensitive to static electricity. Wear an anti-static wrist strap connected to earth ground. Place the instrument top cover (connected to earth ground) near the work area with the bare metal side facing up.
85103E Upgrade Package (8510A to 8510C) Rerouting Cables Rerouting Cables Use the following procedure to install an SMB adapter tee, and to reroute cables in the 85102. 1. Refer to Figure 7. On the A23 assembly, install the supplied SMB adapter tee on A23J1.
85103E Upgrade Package (8510A to 8510C) Attaching New Labels Attaching New Labels Procedure There are two grey labels in the upgrade kit, but you will use only one depending on the front dress panel of the analyzer (check the label size). The rest of the labels are metallic silver.
Page 31
85103E Upgrade Package (8510A to 8510C) Attaching New Labels Figure 8 Location of New Labels If you do not use the supplied optional serial tags to note the original NOTE 85101 and 85102 serial numbers, dispose of the unused tags.
85103E Upgrade Package (8510A to 8510C) Upgrading an 8360 Source Upgrading an 8360 Source Refer to the chapter on “Main Troubleshooting Procedure” in the 8510C Network Analyzer On-Site Service Manual (p/n 08510-90282). After upgrading source(s), continue with “Reconnecting the System.”...
85103E Upgrade Package (8510A to 8510C) Reconnecting the System Reconnecting the System Procedure Use the following procedure to connect the 8510C system: 1. Using the 85101C, and the newly upgraded 85102C, connect the system as shown in Figure 2. Continue with “Loading the 8510C Operating System.”...
85103E Upgrade Package (8510A to 8510C) Loading the 8510C Operating System Loading the 8510C Operating System After you load the operating system, you will have an 8510C network analyzer, and it will be referenced as such. Often the letter designation “C” at the end of the 8510C model number is omitted.
Does the analyzer display any internal self-test failure on the self-test status LEDs or display? If not, the instrument passed the internal self tests, and functions fully. If a self-test fails, or a running error is displayed, refer to the Agilent Technologies 8510C On-Site Service Manual.
85103E Upgrade Package (8510A to 8510C) Checking the 8510C LO Phase Lock Out Functional Check The A16 assembly inverts the main phase lock signal to fine tune a sweep oscillator. In this check, you use the 8510C as an oscilloscope to check this function.
Page 37
85103E Upgrade Package (8510A to 8510C) Checking the 8510C 5. Turn off the 8510. 6. On the 85102: a. Remove the top cover. b. Disconnect W24 from the A17J4 Ref X connector. c. Connect the BNC-to-SMB cable provided in the upgrade package from the rear panel LO Phase Lock Out jack to the A17J4 Ref X connector.
85103E Upgrade Package (8510A to 8510C) Checking the 8510C If the trace flattens out and the display shows an IF Failed Cal NOTE message, press to clear the error. This is not a true error. ENTRY OFF 14. Continue with “Trigger In Functional Check.”...
Page 39
85103E Upgrade Package (8510A to 8510C) Checking the 8510C Figure 12 Typical Partial Trace Confirming Trigger In Circuit Operation 6. Turn off the 8510. 7. On the 85102: a. Remove the SMB-to-BNC cable. b. For an 8360 series source, reconnect the Trigger In cable.
If the 8510A was equipped with the time domain option (Option 010), either originally or through an upgrade, then the 85103E upgrade package ordered should also have been equipped with Option 010. If so, there should be a serial number label on the back of the 85101C with OPT: 010 printed on it.
85103E Upgrade Package (8510A to 8510C) Checking Calibration Kit Definitions Checking Calibration Kit Definitions After an upgrade, an 8510 contains calibration kit definitions that may be significantly different from those that were in the pre-upgraded analyzer. Ensure that the calibration kit definitions contained in the 8510 match the calibration kit used by reloading the definitions from disk.