Making Measurements
Example 7: Stimulus-Response Measurements
Figure 2-38
Measure the rejection range with delta markers.
Activating normalization changes the softkeys that appear in the
amplitude menu:
appears, and
is replaced by
RANGE LVL
REF LVL
NORM
L. Although both these functions reposition the trace on the
REF LV
display,
adjusts attenuation and gain, while
RANGE LVL
NORM REF LVL
does not change the spectrum analyzer settings.
, like
, is a trace math function that lets
NORM REF LVL
NORM REF POSN
you change the position of the normalized trace within the range of the
graticule, without changing spectrum analyzer settings. See Figure
2-39. As a result, your measurement remains calibrated. However,
using these two trace math functions may mean that the top of the
screen does not represent the gain-compression limit.
92
Chapter 2