1. Press
to turn the analyzer on.
LINE
2. The analyzer takes about half a minute to perform a series of
self-diagnostic and adjustment routines. At completion, the screen
displays the analyzer model number and the firmware date (for
example, 890802 indicates August 2, 1989). If you should ever need
to call Agilent Technologies for service or have any questions
regarding your analyzer, it is helpful to know the firmware date to
get the most accurate information.
3. Allow the analyzer a 5-minute warmup. Refer to the warmup
specification in the calibration guide.
4. The reference level calibration adjustment procedure in this chapter
should be performed before checking the specifications.
5. The trace alignment procedure can be performed to optimize the
display location.
6. The GPIB address may need to be changed for remote operation.
Chapter 1
Turning the Spectrum Analyzer On for the First Time
Quick Start Guide