Agilent Technologies 8560E User Manual page 704

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

If You Have a Problem
Error Messages
SYSTEM Hardware/firmware interaction, floating
ERR 751
overflow; check other errors.
SYSTEM Hardware/firmware interaction, floating
ERR 752
underflow; check other errors.
SYSTEM Hardware/firmware interaction, LOG error;
ERR 753
check other errors.
SYSTEM Hardware/firmware interaction, Integer overflow;
ERR 754
check other errors.
SYSTEM Hardware/firmware interaction, squareroot error;
ERR 755
check other errors.
SYSTEM Hardware/firmware interaction, Triple overflow;
ERR 756
check other errors.
SYSTEM Hardware/firmware interaction, BCDL overflow;
ERR 757
check other errors.
SYSTEM Unknown system error.
ERR 758
SYSTEM Hardware/firmware interaction. Code invoked for
ERR 759
wrong instrument.
NO FADC The power-on read of the fast analog to digital
ERR 760
converter hardware is failing.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
Error codes 800 through 899, MODULE, are reserved for option
modules, such as the Agilent 85629B test and adjustment module or
the Agilent 85620A mass memory module. Refer to the option module
manual for a list of product-specific error messages.
Error codes 900 to 910 relate to user-generated errors. These
errors occur if the operator has entered information incorrectly, or is
attempting to use the analyzer inappropriately.
TG UNLVL Tracking generator output is unleveled.
ERR 900
TGFrqLmt Tracking generator output unleveled because
ERR 901
START FREQ is set below tracking generator
frequency limit (300 kHz).