Agilent Technologies 8560E User Manual page 702

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

If You Have a Problem
Error Messages
RBW 300K Unable to adjust 300 kHz RES BW.
ERR 585
RBW 1M Unable to adjust 1 MHz RES BW.
ERR 586
RBW 30K Unable to adjust 30 kHz RES BW.
ERR 587
RBW 100K Unable to adjust 100 kHz RES BW.
ERR 588
RBW 300K Unable to adjust 300 kHz RES BW.
ERR 589
RBW 1M Unable to adjust 1 MHz RES BW.
ERR 590
LOG AMPL Unable to adjust amplitude in log scale.
ERR 591
LOG AMPL Unable to adjust amplitude in log scale.
ERR 592
LOG TUNE LIMIT Cal tune error from DC logger
ERR 593
calibration.
LOG OFST Attenuator calibration offset error from DC
ERR 594
logger calibration.
LOG ATTN Attenuator calibration absolute error from DC
ERR 595
logger calibration.
LOG FID Fidelity error from DC logger calibration.
ERR 596
LOG OFST Fidelity offset error from DC logger
ERR 597
calibration.
LOG OFST Fidelity offset unstable from DC logger
ERR 598
calibration.
LOG GAIN Fidelity gain error from DC logger calibration.
ERR 599
Error codes 600 through 699 relate to system failures.
Instrument service is required.
SYSTEM Hardware/firmware interaction; check other
ERR 600
errors.
SYSTEM Hardware/firmware interaction; check other
ERR 601
errors.
OUTOF RG ADC input is outside of the ADC range.
ERR 650
NO IRQ Microprocessor is not receiving interrupt from
ERR 651
ADC.