Internal Preamp

Accesses a menu of keys that control the internal preamps. Turning on the preamp gives a better noise
figure, but a reduced TOI to noise floor dynamic range. You can optimize this setting for your particular
measurement. The LowBand selection needs to show "(3.0 GHz)" for all versions of the N9000A and
"(3.6 GHz)" for the other models.
See
"Internal Preamp" on page
This is only available when the selected input is RF.
Key Path
Initial S/W Revision

Ref Position

Allows you to set the reference position to either Top, Ctr (center) or Bottom.
Ref Position (Spectrum)
Allows you to set the spectrum reference position to Top, Ctr (center) or Bottom.
Key Path
Mode
Remote Command
Example
Notes
Preset
State Saved
Range
Initial S/W Revision
Ref Position (IQ Waveform)
Allows you to set the spectrum reference position to Top, Ctr (center) or Bottom.
Key Path
Mode
381for more information.
AMPTD Y Scale
Prior to A.02.00
AMPTD Y Scale
BASIC
:DISPlay:SPECtrum:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RPO
Sition TOP|CENTer|BOTTom
:DISPlay:SPECtrum:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RPO
Sition?
DISP:SPEC:VIEW:WIND2:TRAC:Y:RPOS CENT
DISP:SPEC:VIEW:WIND2:TRAC:Y:RPOS?
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
TOP
Saved in instrument state.
Top | Ctr | Bot
Prior to A.02.00
AMPTD Y Scale
BASIC
Complex Spectrum Measurement
AMPTD Y Scale
223