Marker To; Mkr -> Cf; Mkr -> Ref Lvl - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Marker To

Accesses menu keys that can copy the current marker value into another instrument parameter (for
example, Center Freq). If the currently selected marker is not on when you press this front panel key, it is
turned on at the center of the screen as a normal marker.
Key Path
Initial S/W Revision
Mkr -> CF
Sets the center frequency to the frequency of the selected marker. The marker stays at this frequency, so
it moves to the center of the display.
Key Path
Mode
Remote Command
Example
Notes
Dependencies
Initial S/W Revision
Mkr -> Ref Lvl
Sets the reference level to the amplitude value of the selected marker, moving the marked point to the
reference level (top line of the graticule).
Key Path
Mode
Remote Command
Example
Front-panel key
Prior to A.02.00
Marker ->
BASIC
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[
:SET]:CENTer
CALC:SPEC:MARK4:CENT
In the delta marker mode, this function sets the center frequency to the x-axis
value of the delta marker. If the currently selected marker is not on when this
key is pressed, it is turned on at the center of the screen as a normal type
marker.
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
This key is not available (grayed out) when the selected marker is not on the
spectrum trace.
Prior to A.02.00
Marker ->
BASIC
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[
:SET]:RLEVel
CALC:SPEC:MARK4:RLEV
Complex Spectrum Measurement
Marker To
239