Notes
Preset
State Saved
Range
Initial S/W Revision
Modified at S/W Revision

Band Adjust

Accesses a menu that enables you to set the frequency span width and the left and right edge, or time
values, for the band or interval of the selected marker.
Key Path
Initial S/W Revision
Band/Interval Span for Time Domain
Sets the width of the frequency span for the selected marker.
Key Path
Mode
Remote Command
Example
Notes
Couplings
Preset
Preset
State Saved
Min
Max
You must be in a mode that includes the Waveform measurement to use this
command. Use INSTrument:SELect to set the mode.
OFF
Saved in instrument state.
Band/Interval Power|Band Interval Density|Marker Function Off
Prior to A.02.00
A.03.00
Marker Function
Prior to A.02.00
Marker Function
BASIC, PN, WCDMA, C2K, GSM, WIMAX OFDMA, TD-SCDMA,
1xEV-DO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE, LTETDD,
Digital Cable TV
:CALCulate:WAVeform:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
FUNCtion:BAND:SPAN <time>
:CALCulate:WAVeform:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
FUNCtion:BAND:SPAN?
CALC:WAV:MARK:FUNC:BAND:SPAN 20 ms
CALC:WAV:MARK:FUNC:BAND:SPAN?
You must be in a mode that includes the Waveform measurement to use this
command. Use INSTrument:SELect to set the mode.
Changing the Band/Interval Span necessarily changes the Band/Interval Left
and Band/Interval Right values
0
10% of Meas Time
Saved in instrument state.
0
100 s