Waveform Measurement

7
The waveform measurement is a generic measurement for viewing the input signal waveforms in the
time domain. This measurement represents how the instrument performs the zero span functionality
found in traditional spectrum analyzers. For more details, see
on page
280" below.
This topic contains the following sections:
"Measurement Commands for Waveform" on page 279
"Remote Command Results for the Waveform Measurement" on page 279
Measurement Commands for Waveform
The general functionality of CONFigure, INITiate, FETCh, MEASure, and READ are described at this
section.
:CONFigure:WAVeform
:CONFigure:WAVeform:NDEFault
:INITiate:WAVeform
:FETCh:WAVeform[n]
:MEASure:WAVeform[n]
:READ:WAVeform[n]
For more measurement related commands, see the SENSe subsystem, and the section
Measurement Functions" on page
Remote Command Results for the Waveform Measurement
The following table denotes the returned results from the FETCh|MEASure|READ commands:
n
Results Returned
0
Returns unprocessed I/Q trace data, as a series of trace point values, in volts. The I values are listed
first in each pair, using the 0 through even-indexed values. The Q values are the odd-indexed values.
483.
""Waveform Measurement Description"
"Remote
279