Restart

The Restart function restarts the current sweep, or measurement, or set of averaged/held sweeps or
measurements. If you are Paused, pressing Restart does a Resume.
The Restart function is accessed in several ways:
• Pressing the Restart key
• Sending the remote command INIT:IMMediate
• Sending the remote command INIT:RESTart
Key Path
Remote Command
Example
Notes
Couplings
Status Bits/OPC dependencies
Initial S/W Revision
Remote Command
Example
Notes
Couplings
Front panel key
:INITiate[:IMMediate]
:INIT:IMM
:INITiate:RESTart
:INITiate:IMMediate
Either of the above commands perform exactly the same function.
Resets average/hold count k. For the first sweep overwrites all active
(update=on) traces with new current data. For application modes, it resets
other parameters as required by the measurement.
This is an Overlapped command.
The STATus:OPERation register bits 0 through 8 are cleared.
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.
The SWEEPING bit is set.
The MEASURING bit is set.
Prior to A.02.00
:INITiate:RESTart
:INIT:REST
:INITiate:RESTart
:INITiate:IMMediate
Either of the above commands perform exactly the same function.
Resets average/hold count k. For the first sweep overwrites all active
(update=on) traces with new current data. For application modes, it resets
other parameters as required by the measurement.
Restart
525