IF Gain Low + IF Gain Offset +4 dB = Total IF Gain of +4 dB (0 + 4 = 4)
IF Gain High + IF Gain Offset +4 dB = Total IF Gain of +14 dB (10 + 4 = 14)
IF Gain Low + IF Gain Offset –6 dB = Total IF Gain of –6 dB (0 – 6 = –6)
IF Gain High + IF Gain Offset –6 dB = Total IF Gain of +6 dB (10 – 6 = 4)
Thus the total IF Gain range when IF Gain Offset is available is a minimum of 0 – 6 = –6 dB and a
maximum of 10 + 6 = 16 dB. The available IF Gain depends on the IF Path and center frequency. The
maximum IF Gain may not be achievable at all times depending on the configuration.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Min
Max
Default Unit

Meas Preset

Returns all measurement local parameters to the factory default values.
Key Path
Mode
Remote Command
Example
Notes
Initial S/W Revision
Meas Setup, Advanced
[:SENSe]:SPECtrum:IF:GAIN:OFFSet <rel_ampl>
[:SENSe]:SPECtrum:IF:GAIN:OFFSet?
SPEC:IF:GAIN:OFFS 2
Sets the IF Gain offset to 2
If the IF Path Selection is B10M or B25M, then this feature is not available
and is grayed out.
0
Saved in instrument state
–6
+6
dB
Meas Setup
BASIC
:CONFigure:SPECtrum
CONF:SPEC
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
Prior to A.02.00
Complex Spectrum Measurement
Meas Setup
259