Complex Spectrum Measurement
BW
BW
Opens the Bandwidth menu.
Key Path
Initial S/W Revision
Res BW
Enables you to set the resolution bandwidth setting. This is the resolution bandwidth of the FFT analysis.
Changing this value changes the FFT Window size, FFT length and the sweep time (measurement
capture length).
If FFT Length Ctrl in the FFT Size menu under Meas Setup, Advanced is set to Manual, the Res BW key
is grayed out and shows the resolution bandwidth determined by the FFT Window size.
If the function is auto-coupled, the value setting is ignored.
Key Path
Mode
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Backwards Compatibility SCPI
Initial S/W Revision
226
Front-panel key
Prior to A.02.00
BW
BASIC
[:SENSe]:SPECtrum:BANDwidth[:RESolution] <bandwidth>
[:SENSe]:SPECtrum:BANDwidth[:RESolution]?
[:SENSe]:SPECtrum:BANDwidth[:RESolution]:AUTO
ON|OFF|1|0
[:SENSe]:SPECtrum:BANDwidth[:RESolution]:AUTO?
SPEC:BAND 100
SPEC:BAND?
SPEC:BAND:AUTO OFF
SPEC:BAND:AUTO?
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
160 kHz
ON
Saved in instrument state.
0.1 Hz
3.0 MHz
[:SENSe]:SPECtrum:BWIDth[:RESolution]
Prior to A.02.00