System Functions
System
If this control is entered without an external keyboard or mouse connected, you can cancel the control by
pressing the Cancel (ESC) front-panel key.
Key Path
Mode
Initial S/W Revision
LXI
Opens a menu that allows you to access the various LXI configuration properties.
Key Path
Initial S/W Revision
LAN Reset
Resets the LAN connection.
Key Path
Initial S/W Revision
System IDN Response
This key allows you to specify a response to the *IDN? query, or to return the analyzer to the Factory
response if you have changed it.
To choose the factory-set response, press the
To specify your own response, press the
Key Path
Mode
Remote Command
Notes
Preset
State Saved
192
System, I/O Config
All
Prior to A.02.00
System, I/O Config
Prior to A.02.00
System, I/O Config, LXI
Prior to A.02.00
key.
Factory
key, and enter your desired response.
User
System, I/O Config
All
:SYSTem:IDN <string>
:SYSTem:IDN?
This affects the response given in all Modes of the Analyzer, unless the
current Mode has also specified a custom response, in which case the
current Mode's custom IDN response takes precedence over the System's,
but only while that Mode is the current Mode.
It survives shutdown and restart of the software and therefore survives a
power cycle
Null string as parameter restores the Factory setting
This is unaffected by Preset but is set to the original factory setting on a
"Restore System Defaults->Misc"
No