Trigger Source Presets
Here are the Trigger Source Presets for the various measurements:
Meas
Mode
Swept SA
SA
CHP
SA, WCDMA,
C2K, WIMAX
OFDMA,
1xEVDO,
DVB-T/H,
DTMB, LTE,
LTETDD,
CMMB, ISDB-T,
Digital Cable TV
OBW
SA, WCDMA,
C2K, WIMAX
OFDMA,
TD-SCDMA,
1xEVDO, LTE,
LTETDD,
CMMB, ISDB-T
CCDF
SA, WCDMA,
C2K,
WIMAXOFDMA
, TD-SCDMA,
1xEV-DO,
DVB-T/H,
DTMB, LTE,
LTETDD,
CMMB, ISDB-T,
Digital Cable TV
Preset for RF
Preset for IQ
IMM
IQ not supported
IMM
IQ not supported
1xEVDO: EXT1
IQ not supported
others: IMM
WIMAXOFDMA
TD-SCDMA and
: RFBurst
1xEV-DO:
BTS: External 1
LTETDD:
MS: IQMag
BTS: External 1
MS: Periodic
LTETDD:
Timer
BTS: External 1
MS: Periodic
TD-SCDMA and
Timer
1xEV-DO:
BTS: External 1
Others: IMM
MS: RFBurst
SA, WCDMA,
C2K, LTE,
CMMB, ISDB-T,
DVB-T/H,
DTMB, Digital
Cable TV:
IMMediate
Notes
For 1xEVDO mode, the trigger
source is coupled with the gate
state, as well as the gate source.
When the trigger source changes to
RFBurst, External1 or External2,
the gate state is set to on, and the
gate source is set identically with
the trigger source. When the trigger
source changes to IMMediate,
VIDeo, LINE, FRAMe or IF, the
gate state is set to off.
For TD-SCDMA:
Trigger source is coupled with radio
device. When radio device changes
to BTS, trigger source will be
changed to EXTernal1. When radio
device changes to MS, trigger
source will be set as RFBurst for
RF or IQ Mag for BBIQ.
When TriggerSource is RFBurst or
IQ Mag, Measure Interval is grayed
out.
Trigger