Input/Output
Initial S/W Revision
Q Port
The Q port calibration is performed with the probe body attached to the front panel's Q port and the
probe tip connected via an adapter to the Cal Out port. The guided calibration will show a diagram of the
required connections.
Show Adapter
Show a connection diagram and instructions for the probe and adapter. See
Key Path
Notes
Initial S/W Revision
Back
Return to the prior step in the calibration procedure.
Key Path
Initial S/W Revision
Next
Perform the Q port calibration.
Key Path
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration
data.
Key Path
480
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Either a passive or an active probe adapter diagram will be shown, depending
on the type of probe attached.
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
:CALibration:IQ:PROBe:Q
CAL:IQ:PROB:Q
The Q port must be connected to the Cal Out port before issuing the SCPI
command.
The calibration data is saved as soon as the port is calibrated and will survive
power cycles. It is not reset by any preset or restore data commands.
No
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
"Show Adapter Screen" on page
482.