Readback
Initial S/W Revision
Off
Turns the display backlight off.
Key Path
Example
Readback
Initial S/W Revision
Backlight Intensity
An active function used to set the backlight intensity. It goes from 0 to 100 where 100 is full on and 0 is
off. This value is independent of the values set under the Backlight on/off key.
Key Path
Remote Command
Example
Preset
Min
Max
Initial S/W Revision

Full Screen

When
Full Screen
is pressed the measurement window expands horizontally over the entire instrument
display. The screen graticule area expands to fill the available display area.
It turns off the display of the softkey labels, however the menus and active functions still work. (Though
it would obviously be very hard to navigate without the key labels displayed.) Pressing
again while Full Screen is in effect cancels Full Screen.
Note that the banner and status lines are unaffected. You can get even more screen area for your data
display by turning off the Meas Bar (in the Display menu) which also turns off the settings panel.
Full Screen is a Meas Global function. Therefore it is cancelled by the
Key Path
Remote Command
On
Prior to A.02.00
View/Display, Display, System Display Settings, Backlight
DISP:BACK OFF
Off
Prior to A.02.00
View/Display, Display, System Display Settings
:DISPlay:BACKlight:INTensity <integer>
:DISPlay:BACKlight:INTensity?
DISP:BACK:INT 50
100 (Set by Restore Misc Defaults)
0
100
Prior to A.02.00
Display
:DISPlay:FSCReen[:STATe] OFF|ON|0|1
:DISPlay:FSCReen[:STATe]?
View/Display
Full Screen
key.
Preset
621