Visual annotation in the Show Alignment Statistics screen
All
:CALibration:TEMPerature:LRF?
:CAL:TEMP:LRF?
Value is in degrees Centigrade at which the last successful Align Now, RF was
executed, either individually or as a component of Align Now, All.
No
Prior to A.02.00
Visual annotation in the Show Alignment Statistics screen
All
:CALibration:TIME:LPReselector?
:CAL:TIME:LPR?
Value is date and time the last successful Characterize Preselector was
executed. The date is separated from the time by a space character. Returns ""
if no Characterize Preselector has ever been performed on the instrument.
In models that do not include preselectors, this command is not enabled and
any attempt to set or query will yield an error.
No
Prior to A.02.00
Visual annotation in the Show Alignment Statistics screen
All
:CALibration:TEMPerature:LPReselector?
:CAL:TEMP:LPR?
Value is in degrees Centigrade at which the last successful Characterize
Preselector was executed.
In models that do not include preselectors, this command is not enabled and
any attempt to set or query will yield an error.
No
Prior to A.02.00
Visual annotation in the Show Alignment Statistics screen
All
:CALibration:AUTO:TIME:OFF?