BW
Accesses a menu that enables you to control the information bandwidth functions of the instrument. You
can also select the filter type for the measurement.
Key Path
Initial S/W Revision

Digital IF BW

Enables you to set the Digital IF (formerly Info BW) bandwidth of the instrument.
Key Path
Mode
Remote Command
Example
Notes
Remote Command Notes
Dependencies
Front-panel key
Prior to A.02.00
BW
BASIC, PN, WCDMA, C2K, GSM, WIMAX OFDMA, TD-SCDMA,
1xEV-DO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE, LTETDD,
Digital Cable TV
[:SENSe]:WAVeform:DIF:BANDwidth <freq>
[:SENSe]:WAVeform:DIF:BANDwidth?
WAV:DIF:BAND 1kHz
WAV:DIF:BAND?
Max value depends on the IF Path Selection
You must be in a mode that includes the Waveform measurements to use this
command. Use INSTrument:SELect to set the mode.
For applications that have the IF Path Selection menu such as the BASIC
mode, if IF Path Auto is OFF, the maximum value depends on which IF Path
is currently selected. If 10 MHz, 25 MHz, 40 MHz, or 140 MHz paths are
selected, the maximum value of this parameter will be 10, 25, 40, or 140
MHz, respectively. If IF Path Auto is ON, the maximum value will be the
maximum Digital IF BW available in the instrument regardless of the current
IF Path Selection. For example, if the instrument had the options B25, B40,
and B1X installed, the maximum available Digital IF BW of the instrument is
140 MHz. Thus, if IF Path Auto is ON and IF Path Selection is 25 MHz, the
maximum Digital IF BW is not limited to 25 MHz but is 140 MHz.