Rf Input - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Remote Command
Example
Notes
Preset
Initial S/W Revision
Input/Output variables - Preset behavior
Virtually all the input/output settings are NOT a part of mode preset. They can be set to their default
value by one of the three ways - by using the Restore Input/Output Defaults key on the first page of the
input/output menu, by using the System->Restore System Defaults->Input/Output Settings or by using
the System -> Restore System Defaults->All. Also, they survive a Preset and a Power cycle.
A very few of the Input/Output settings do respond to a Mode Preset; for example, if the Calibrator is on
it turns off on a Preset, and if DC coupling is in effect it switches to AC on a Preset. These exceptions are
made in the interest of reliability and usability, which overrides the need for absolute consistency.
Exceptions are noted in the SCPI table for the excepted functions.

RF Input

Selects the front-panel RF input port to be the analyzer signal input. If RF is already selected, pressing
this key accesses the RF input setup functions.
Key Path
Example
Readback
Initial S/W Revision
394
:INPut:MIXer EXTernal|INTernal
:INPut:MIXer?
INP:MIX INT
INP:MIX?
INPut:MIXer EXTernal|INTernal legacy command is mapped as follows:
1. When INPut:MIXer EXTernal is received, SENSe:FEED EMIXer is
executed.
2. When INPut:MIXer INTernal is received, SENSe:FEED RF is executed.
3. When INPut:MIXer? is received, the response will be INT if any input
other than the external mixer is selected and EXT if the external mixer is
selected
INT
A.08.01
Input/Output
[:SENSe]:FEED RF
The RF input port, RF coupling, and current input impedance settings appear
on this key as:
"XX, YY, ZZ" where
XX is RF, RF2, RFIO1, RFIO2, depending on what input is selected (only
appears on analyzers with multiple RF inputs)
YY is AC or DC
ZZ is 50 or 75
Prior to A.02.00

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9020aN9010aN9000aN9038a

Table of Contents