Auto Scaling

Toggles the Auto Scaling function between On and Off. When the
this function automatically determines the scale per division and reference values based on the
measurement results.
Key Path
Mode
Remote Command
Example
Notes
Couplings
Preset
State Saved
Range
Initial S/W Revision
Modified at S/W Revision
AMPTD Y Scale
BASIC, PN, WCDMA, C2K, GSM, WIMAX OFDMA, TD-SCDMA,
1xEV-DO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE, LTETDD,
Digital Cable TV
:DISPlay:WAVeform:VIEW[1]|2:WINDow[1]:TRACe:Y[:SCALe]:C
OUPle 0|1|OFF|ON
:DISPlay:WAVeform:VIEW[1]|2:WINDow[1]:TRACe:Y[:SCALe]:C
OUPle?
DISP:WAV:VIEW:WIND:TRAC:Y:COUP OFF
DISP:WAV:VIEW:WIND:TRAC:Y:COUP?
You must be in a mode that includes the Waveform measurement to use this
command. Use INSTrument:SELect to set the mode.
When Auto Scaling is On, upon pressing the Restart front-panel key, this
function automatically switches the scale per division and reference values
into the defaults.
When you set a value to either Scale/Div or Ref Value manually, Auto Scaling
automatically changes to Off.
OFF
Saved in instrument state.
On|Off
Prior to A.02.00
A.03.00
Waveform Measurement
AMPTD Y Scale
Restart
front-panel key is pressed,
287