System Functions
System
Initial S/W Revision

Power On

Enables you to select how the instrument should power on. The options are: Mode and Input/Output
Defaults, User Preset and Last State.
Key Path
Mode
Remote Command
Example
Preset
Preset
State Saved
Initial S/W Revision
Mode and Input/Output Defaults
When the analyzer is powered on in Mode and Input/Output Defaults, it performs a Restore Mode
Defaults to all modes in the instrument and also performs a Restore Input/Output Defaults.
Persistent parameters (such as Amplitude Correction tables or Limit tables) are not affected at poweron,
even though they are normally cleared by Restore Input/Output Defaults and/or Restore Mode Defaults.
Key Path
Mode
Example
Readback Text
Initial S/W Revision
User Preset
Sets
to
Power On
User Preset
mode and switch to the power-on mode. Power On User Preset will not affect any settings beyond what a
normal User Preset affects.
An instrument could never power up for the first time in User Preset.
NOTE
Key Path
152
Prior to A.02.00
System
All
:SYSTem:PON:TYPE MODE|USER|LAST|PRESet
:SYSTem:PON:TYPE?
:SYST:PON:TYPE MODE
MODE
This is unaffected by a Preset but is set to Mode on a "Restore System
Defaults->All"
No
Prior to A.02.00
System, Power On
All
SYST:PON:TYPE MODE
Defaults
Prior to A.02.00
. When the analyzer is powered on in User Preset, it will User Preset each
System, Power On