Restore Input/Output Defaults; Data Source - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output

Restore Input/Output Defaults

This selection causes the group of settings and data associated with the
their default values. In addition, when a Source is installed, licensed and selected, Restore Input/Output
defaults will initiate a Source Preset.
This level of Restore System Defaults does not affect any other system settings or mode settings and
does not cause a mode switch. All the features described in this section are reset using this key, including
Input Corrections and Data (described in the Corrections section).
Key Path
Example
Notes
Initial S/W Revision

Data Source

Gives you the choice of either using a hardware input signal as the input or raw data stored in a data
storage buffer from an earlier acquisition. You can also share raw data across certain measurements that
support this feature. The measurements must be capable of storing raw data. There are three choices
under this menu. You can select "Inputs" which is the same as selecting one of the inputs from the input
port, for example RF, AREF, I/Q, or IFALign. Selecting "Capture Buffer" allows you to use data that has
been stored earlier in the same measurement or from a previous measurement using the "Current Meas
-> Capture Buffer" feature. Selecting "Recorded Data" allows you to playback long data capture records
stored in the record buffer.
When you make a recording (see
the Recall section) the data source is automatically set to Recorded Data. You can toggle the data source
between Inputs and the current Recording (if there is one). That is, the recording remains in memory
until it is replaced by a new recording, or the application is closed.
Key Path
Remote Command
Example
Notes
Dependencies
436
Input/Output
:SYST:DEF INP presets all the Input/Output variables to their factory default
values.
Refer to the Utility Functions for information about Restore System Defaults
and the complete description of the :SYSTem:DEFault INPut: command.
Prior to A.02.00
"Record Data Now" on page
Input/Output
[:SENSe]:FEED:DATA INPut|STORed|RECorded
[:SENSe]:FEED:DATA?
FEED:DATA REC
FEED:DATA?
INPuts = Inputs
STORed = Capture Buffer
RECorded = Record Data Buffer
Not all inputs are available in all modes. Unavailable keys are grayed out.
Input/Output
key to be a reset to
438) or when you recall a recording (see

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9020aN9010aN9000aN9038a

Table of Contents