Auto Scaling - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

State Saved
Range
Initial S/W Revision
Modified at S/W Revision

Auto Scaling

Toggles the scale coupling function between On and Off.
Key Path
Mode
Remote Command
Example
Notes
Couplings
Preset
State Saved
Range
Initial S/W Revision
Modified at S/W Revision
Saved in instrument state.
Left|Ctr|Right
Prior to A.02.00
A.03.00
SPAN X Scale
BASIC, PN, WCDMA, C2K, GSM, WIMAX OFDMA, TD-SCDMA,
1xEV-DO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE, LTETDD,
Digital Cable TV
:DISPlay:WAVeform:VIEW[1]|2:WINDow[1]:TRACe:X[:SCALe]:C
OUPle 0|1|OFF|ON
:DISPlay:WAVeform:VIEW[1]|2:WINDow[1]:TRACe:X[:SCALe]:C
OUPle?
DISP:WAV:VIEW:WIND:TRAC:X:COUP ON
DISP:WAV:VIEW:WIND:TRAC:X:COUP?
You must be in a mode that includes the Waveform measurement to use this
command. Use INSTrument:SELect to set the mode.
When Auto Scaling is On and the Restart front-panel key is pressed, this
function automatically determines the scale per division and reference values
based on the measurement results.
When you set a value to either Scale/Div or Ref Value manually, Auto Scaling
automatically changes to Off.
1
Saved in instrument state.
On|Off
Prior to A.02.00
A.03.00
Waveform Measurement
Span X Scale