Running the Calibrations (LO, RF, and IF)

These 3 Cals should be run in the order shown:
1.
2.
3.

LO Calibration

1. Before starting the LO calibration disconnect the (Test Set to PNA) rear
2. In the Banded Installation CAL window, click "Begin LO".
Keysight Self-Configured System Installation Guide
"LO Calibration" on page 3-23
"RF Calibration" on page 3-26
"IF Calibration" on page 3-28
panel LO cable from the PNA. The Banded LO Cal requires a 3.5 mm Power
Sensor to be connected to the PNA rear panel LO connector.
The PNA LO Banded Adjustment window opens. See
page
24.
Figure 3-17 on
3-23