Which Cleaning Procedure should I use?

How to clean connectors

Keysight N77xx Series
cause damage to sensitive components, and in addition your warranty
will be voided.
Never open the instruments as they can be damaged. Opening the
instruments puts you in danger of receiving an electrical shock from
your device, and renders your warranty void.
Light dirt
If you just want to clean away light dirt, observe the following procedure
for all devices:
1 Use compressed air to blow away large particles.
2 Clean the device with a dry cotton swab.
3 Use compressed air to blow away any remaining filament left by the
swab.
Heavy dirt
If the above procedure is not enough to clean your instrument, follow
one of the procedures below. Please consult
this Instrument" on page 84
instrument.
If you are unsure of how sensitive your device is to cleaning, please
contact the manufacturer or your sales distributor
Cleaning connectors is difficult as the core diameter of a single-mode
fiber is only about 9 µm. This generally means you cannot see streaks
or scratches on the surface. To be certain of the condition of the surface
of your connector and to check it after cleaning, you need a microscope.
In the case of scratches, or of dust that has been burnt onto the surface
of the connector, you may have no option but to polish the connector.
This depends on the degree of dirtiness, or the depth of the scratches.
This is a difficult procedure and should only be performed by skilled
personal, and as a last resort as it wears out your connector.