Performance Tests and Checks
System Performance Verification

Verification Procedure

1. Insert the USB memory stick from the 85059A kit into the PNA.
2. Press Utility > System > Service > System Verification. The System Verification dialog
box is displayed; refer to
Figure 6-3
System Verification Dialog Box
3. In the Select Calibration Kit drop-down menu, select the "Agilent 85059A Data-based (1.00
mm)" calibration kit by clicking on it. The corresponding verification kit to use is automatically
selected. Refer to
Figure
4. Under Test Cable(s):
• Select Single Cable (+ Adapters)
• Set C able Flex Factor to 100%
5. Under Outputs, the recommended selections are "File Tabular Data" and "File Graphs." Other
selections may be made. For additional information, click on the Help tab, then click Detailed
Help. This will take you into the PNA Help system. Scroll down to Step 5, "Under Printer
Output...".
6. Select the Configure tab. The window shown in
7. Under Measure Ports, select the checkbox for the desired pair of ports. (The verification
process is performed on pairs of ports.)
8. Under Measure Devices, select both checkboxes.
6-8
Figure
6-3.
6-3.
PNA Series Microwave Network Analyzer System
Figure 6-4 on page 6-9
Installation and Service Guide N5251-90001
N5251A
is displayed.