To restore raw measurement data

1 Access the analyzer's file system as the control user.
Refer to the chapter "Accessing the Logic Analyzer File System".
2 Copy the data.raw file to the appropriate \slot_{x} directory.
For analyzer data, this would be the \slot_a directory. For oscilloscope
data, it would be \slot_b. If you copy the raw data to a different
directory than you originally got it from, the logic analyzer won't know
how to interpret it.
Raw measurement data files are binary format files. Because they are
binary files, if you are using ftp to upload or download them, set the
type to binary by typing "bin" at the ftp prompt.
The data.raw file is overwritten each time new data is acquired.